Comparison of versions 1.0 AND 1.5 of the UltraSensitive AMPLICOR HIV-1 MONITOR test for subjects with low viral load.

@article{Jackson2004ComparisonOV,
  title={Comparison of versions 1.0 AND 1.5 of the UltraSensitive AMPLICOR HIV-1 MONITOR test for subjects with low viral load.},
  author={Jay Brooks Jackson and Estelle Piwowar-Manning and Letanya Johnson-Lewis and Roland Bassett and Lisa M. Demeter and Donald Brambilla},
  journal={Journal of clinical microbiology},
  year={2004},
  volume={42 6},
  pages={2774-6}
}
We compared the performance of two UltraSensitive AMPLICOR HIV-1 MONITOR kits (version 1.5 [v1.5] versus v1.0) by retesting 404 plasma samples with low viral loads (<3,000 copies/ml) with both kits. With 292 samples that initially had <50 copies/ml by the v1.0 kit, the v1.5 assay was more sensitive than the v1.0 assay for samples with human immunodeficiency virus type 1 RNA near the 50-copy/ml cutoff (P = 0.0146). Median numbers of copies per milliliter were similar for 112 samples with 50 to 3… CONTINUE READING