Comparison of the cyclic fatigue resistance of 5 different rotary pathfinding instruments made of conventional nickel-titanium wire, M-wire, and controlled memory wire.

@article{Capar2015ComparisonOT,
  title={Comparison of the cyclic fatigue resistance of 5 different rotary pathfinding instruments made of conventional nickel-titanium wire, M-wire, and controlled memory wire.},
  author={Ismail Capar and M E Kaval and H{\"u}seyin Ertaş and B Hakan Şen},
  journal={Journal of endodontics},
  year={2015},
  volume={41 4},
  pages={535-8}
}
INTRODUCTION This study compared the cyclic fatigue resistance of current nickel-titanium rotary path-finding instruments. METHODS Five types of nickel-titanium rotary pathfinding instruments were used in steel canals with a 90° curvature and a curvature radius of 3 mm (n = 10) and 5 mm (n = 10). The cyclic fatigue of the following instruments was tested at 4 mm from the tip: PathFile (#16 and a .02 taper; Dentsply Maillefer, Ballaigues, Switzerland), G-File (#12 and a .03 taper; Micro-Mega… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 15 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 36 references

Comparison of cyclic fatigue resistance of novel nickel-titanium rotary instruments.

Australian endodontic journal : the journal of the Australian Society of Endodontology Inc • 2015
View 1 Excerpt

Similar Papers