Comparison of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study

@article{Shin2016ComparisonOS,
  title={Comparison of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study},
  author={Hyun Joo Shin and Myung-Joon Kim and Ha Yan Kim and Yun Ho Roh and Mi-Jung Lee},
  journal={European Radiology},
  year={2016},
  volume={26},
  pages={3361-3367}
}
To investigate consistency in shear wave velocities (SWVs) on ultrasound elastography using different machines, transducers and acquisition depths. The SWVs were measured using an elasticity phantom with a Young’s modulus of 16.9 kPa, with three recently introduced ultrasound elastography machines (A, B and C from different vendors) and two transducers (low and high frequencies) at four depths (2, 3, 4 and 5 cm). Mean SWVs from 15 measurements and coefficient of variations (CVs) were compared… CONTINUE READING
Recent Discussions
This paper has been referenced on Twitter 2 times over the past 90 days. VIEW TWEETS

Citations

Publications citing this paper.
Showing 1-10 of 11 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 29 references

Twodimensional shear wave elastography as promising diagnostic tool for predicting malignant thyroid nodules: a prospective singlecentre experience

  • B Liu, J Liang, ZhengY, X Xie, HuangG, L Zhou
  • Eur Radiol
  • 2015

Similar Papers

Loading similar papers…