Comparison of performance across three directional hearing aids.

@article{Ricketts1999ComparisonOP,
  title={Comparison of performance across three directional hearing aids.},
  author={Todd Andrew Ricketts and Sumitrajit Dhar},
  journal={Journal of the American Academy of Audiology},
  year={1999},
  volume={10 4},
  pages={180-9}
}
This study compared the speech recognition performance of 12 hearing-impaired listeners fit with three commercially available behind-the-ear hearing aids in both directional and omnidirectional modes. One digitally programmable analog and two "true digital" hearing aids were selected as test instruments. Testing was completed in both "living room" and anechoic room environments. Speech recognition was examined using modified forms of the Hearing in Noise Test and the Nonsense Syllable Test. The… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 31 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 18 references

Phoneme errors on a nonsense syllable test

  • SB Resnick, Dubno, S Hoffnung, H Levitt
  • JAcoust Soc Am
  • 1975
Highly Influential
13 Excerpts

Real - world performance of an ITE directional microphone

  • M Killion, R Schulein, +4 authors K Chung
  • Hear J
  • 1998

Clinical evaluation of multi - microphone hearing instruments

  • T Voss
  • Hear Rev
  • 1997
1 Excerpt

HINT thresholds for dual - microphone BTE

  • J Agnew, M Block
  • Hear Rev
  • 1997
1 Excerpt

Method for the Calculation of the Speech Intelligibility Index

  • . AmericanNationalStandardsInstitute
  • (ANSI 53.79-1997) . New York : ANSI.
  • 1997

Microphones: Technology and Technique

  • J. Borwick
  • 1997

Predicting Consonant Confusions in Noise on the Basis of Acoustical Analyses

  • JR Dubno
  • Unpublished doctoral dissertation, City…
  • 1997

Similar Papers

Loading similar papers…