Comparison of lesion detection and quantitation of tracer uptake between PET from a simultaneously acquiring whole-body PET/MR hybrid scanner and PET from PET/CT

@article{Wiesmueller2012ComparisonOL,
  title={Comparison of lesion detection and quantitation of tracer uptake between PET from a simultaneously acquiring whole-body PET/MR hybrid scanner and PET from PET/CT},
  author={Marco Wiesmueller and Harald Quick and Bharath K. Navalpakkam and Michael M. Lell and Michael Uder and Philipp Ritt and Daniela Schmidt and Michael Alexander Beck and Torsten Kuwert and Carl C. von Gall},
  journal={European Journal of Nuclear Medicine and Molecular Imaging},
  year={2012},
  volume={40},
  pages={12-21}
}
PET/MR hybrid scanners have recently been introduced, but not yet validated. The aim of this study was to compare the PET components of a PET/CT hybrid system and of a simultaneous whole-body PET/MR hybrid system with regard to reproducibility of lesion detection and quantitation of tracer uptake. A total of 46 patients underwent a whole-body PET/CT scan 1 h after injection and an average of 88 min later a second scan using a hybrid PET/MR system. The radioactive tracers used were 18F… CONTINUE READING
Recent Discussions
This paper has been referenced on Twitter 1 time over the past 90 days. VIEW TWEETS

Citations

Publications citing this paper.
Showing 1-10 of 54 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 41 references

a phantom study

  • S Kumar, AK Pandey, P Sharma, A Malhotra, PET-CT Kumar R. Optimization of the CT acquisition protoc for
  • Nucl Med Commun. 2012;33(2):164–70. 20 Eur J…
  • 2013
1 Excerpt

Similar Papers