Comparison of hip simulator wear of 2 different highly cross-linked ultra high molecular weight polyethylene acetabular components using both 32- and 38-mm femoral heads.

@article{Estok2007ComparisonOH,
  title={Comparison of hip simulator wear of 2 different highly cross-linked ultra high molecular weight polyethylene acetabular components using both 32- and 38-mm femoral heads.},
  author={Daniel M. Estok and Brian R Burroughs and Orhun Kamil Muratoglu and William H. Harris},
  journal={The Journal of arthroplasty},
  year={2007},
  volume={22 4},
  pages={581-9}
}
Much evidence suggests that ultra high molecular weight polyethylene (UHMWP) irradiated to 9.5 Mrad has lower wear than UHMWP given 5 Mrad. Curiously, highly cross-linked polyethylenes gain weight during hip simulator testing. We postulated that (a) UHMWP irradiated to 9.5 Mrad would wear less than UHMWP irradiated to 5 Mrad UHMWP, (b) articulation against 38-mm heads would accentuate this difference, and (c) weight gain of highly cross-linked material reflects the inadequacy of load soak… CONTINUE READING