Comparison of electrical velocimetry and thermodilution techniques for the measurement of cardiac output.

@article{Zoremba2007ComparisonOE,
  title={Comparison of electrical velocimetry and thermodilution techniques for the measurement of cardiac output.},
  author={Norbert Zoremba and Johannes Bickenbach and Beate Krauss and Rolf Rossaint and Ralf Kuhlen and Gereon Schaelte},
  journal={Acta anaesthesiologica Scandinavica},
  year={2007},
  volume={51 10},
  pages={1314-9}
}
AIM To compare a new method of non-invasive determination of cardiac output based on electrical velocimetry (EV-CO) with invasive thermodilution methods. METHODS Fifty critically ill patients were enrolled into the study. EV-CO was compared with cardiac output measured by a pulmonary artery catheter (PA-CO) in one group (n= 25) and by a femoral artery catheter (PiCCO-CO) in a second group (n= 25), by simultaneous measurements. Standard electrocardiography electrodes were used for non-invasive… CONTINUE READING