Comparison of attraction capabilities associated with high-speed, dual-pneumatic vitrectomy probes.

@article{Dugel2015ComparisonOA,
  title={Comparison of attraction capabilities associated with high-speed, dual-pneumatic vitrectomy probes.},
  author={Pravin U. Dugel and Dina Joy K. Abulon and Ramon Dimalanta},
  journal={Retina},
  year={2015},
  volume={35 5},
  pages={915-20}
}
PURPOSE To measure membrane attraction capabilities of enhanced 27-gauge, enhanced 25-gauge, and 23-gauge vitrectomy probes under various parameters. METHODS A membrane-on-cantilever apparatus was used to measure membrane attraction for enhanced 27-, enhanced 25-, and 23-gauge UltraVit probes (n = 6 for each). The following parameters were evaluated: effects of cut rates and duty cycles on membrane attraction distances, and flow rates and vacuum levels required to attract a membrane at a… CONTINUE READING