Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry.

@article{Hoffer2016ComparisonOA,
  title={Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry.},
  author={Kenneth J. Hoffer and Peter Christian Hoffmann and Giacomo Savini},
  journal={Journal of cataract and refractive surgery},
  year={2016},
  volume={42 8},
  pages={
          1165-72
        }
}
PURPOSE To investigate agreement between the IOLMaster 700 based on swept-source optical coherence tomography (SS-OCT) and the Lenstar LS 900 based on optical low-coherence reflectometry (OLCR). SETTING Private practice, Castrop-Rouxelle, Germany. DESIGN Prospective evaluation of diagnostic test. METHODS Axial length (AL), keratometry (K), anterior chamber depth (corneal epithelium to lens) (ACD), lens thickness, and central corneal thickness (CCT) were measured in 183 eyes of 183… CONTINUE READING
BETA

Citations

Publications citing this paper.
SHOWING 1-10 OF 11 CITATIONS

Updates in Biometry.

VIEW 3 EXCERPTS
CITES RESULTS & METHODS
HIGHLY INFLUENCED

References

Publications referenced by this paper.
SHOWING 1-10 OF 26 REFERENCES

ung€ or SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500

A Akman, L Asena
  • Br J Ophthalmol 2015 Dec 16. [Epub ahead of print]
  • 2015
VIEW 3 EXCERPTS
HIGHLY INFLUENTIAL

J.Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China

KJ Hoffer, ShammasHJ, G Savini, Huang
  • J Cataract Refract Surg
  • 2016

Multicenter study of optical low - coherence interferometry and partial - coherence interferometry optical biometers with patients from the United States and China

KJ Hoffer, ShammasHJ, G Savini, J Huang
  • J Cataract Refract Surg
  • 2016

Comparison of 3 biometry devices in cataract patients

S Goebels, M Pattm€ oller, +3 authors A. Langenbucher
  • J Cataract Refract Surg
  • 2015

Evaluation and comparison of the new swept source OCT - based IOLMaster 700 with the IOL - Master 500

WA Lyle, Jin GJC
  • Br J Ophthalmol
  • 2015

Similar Papers