Comparison of MacKay-Marg, Goldmann, and Perkins tonometers in abnormal corneas.

@article{McMillan1975ComparisonOM,
  title={Comparison of MacKay-Marg, Goldmann, and Perkins tonometers in abnormal corneas.},
  author={Franklin D. McMillan and Richard K Forster},
  journal={Archives of ophthalmology},
  year={1975},
  volume={93 6},
  pages={420-4}
}
Intraocular pressure was determined manometrically in owl monkey eyes with normal and edematous corneas. The pressure was measured with Perkins hand-held, Goldmann, and MacKay-Marg tonometers to compare relative accuracy. At monometric levels of 20, 30, and 40 mm Hg, a statistically significant difference in readings was found between normal and edematous corneas with use of the Perkins and Goldmann applanators, whereas no such difference was found with the MacKay-Marg tonometer. The MacKay… CONTINUE READING