Comparison between Tonopachy and other tonometric and pachymetric devices.

@article{Lee2011ComparisonBT,
  title={Comparison between Tonopachy and other tonometric and pachymetric devices.},
  author={Yong Gi Lee and Ji Hyun Kim and Na Rae Kim and Chan Yun Kim and Eun Suk Lee},
  journal={Optometry and vision science : official publication of the American Academy of Optometry},
  year={2011},
  volume={88 7},
  pages={843-9}
}
PURPOSE This study aimed to compare central corneal thickness (CCT) measurement by Tonopachy to that by Pentacam and ultrasound pachymetry, and intraocular pressure (IOP) measurement to that by Goldmann applanation tonometry (GAT). The reproducibility of CCT and IOP measurements by Tonopachy was also evaluated. METHODS In 104 eyes of 104 patients, CCT was… CONTINUE READING