Comparative evaluation of TRI-DOT Rapid HIV test with fourth-generation ELISA for the detection of human immunodeficiency virus.

@article{Sudha2005ComparativeEO,
  title={Comparative evaluation of TRI-DOT Rapid HIV test with fourth-generation ELISA for the detection of human immunodeficiency virus.},
  author={Talasila Sudha and Vijay Dharma Teja and Muralikrishna Gopal and Manoharan Rajesh and V. G. Ishwarya Lakshmi},
  journal={Clinical microbiology and infection : the official publication of the European Society of Clinical Microbiology and Infectious Diseases},
  year={2005},
  volume={11 10},
  pages={
          850-2
        }
}
This study evaluated the TRI-DOT Rapid HIV test for the early detection of human immunodeficiency virus (HIV) infection in comparison with a fourth-generation ELISA (Vironostika HIV Uniform II). Of 23,609 sera, seven (0.03%) gave discordant results. Six of these were reactive only by the fourth-generation assay and were p24 antigen-positive by VIDAS DUO, Western blot and qualitative RT-PCR tests. The remaining discordant serum was considered to be false-positive by the TRI-DOT assay, as it was… CONTINUE READING