Commissioning of a dedicated commercial Co‐60 total body irradiation unit

@inproceedings{Burmeister2018CommissioningOA,
  title={Commissioning of a dedicated commercial Co‐60 total body irradiation unit},
  author={Jay Burmeister and Adrian Nalichowski and Michael G. Snyder and Robert Halford and Geoff Baran and Brian Loughery and Ahmad Hammoud and Joe Rakowski and Todd Bossenberger},
  booktitle={Journal of applied clinical medical physics},
  year={2018}
}
We describe the commissioning of the first dedicated commercial total body irradiation (TBI) unit in clinical operation. The Best Theratronics GammaBeam 500 is a Co-60 teletherapy unit with extended field size and imaging capabilities. Radiation safety, mechanical and imaging systems, and radiation output are characterized. Beam data collection, calibration, and external dosimetric validation are described. All radiation safety and mechanical tests satisfied relevant requirements and measured… CONTINUE READING
Recent Discussions
This paper has been referenced on Twitter 3 times over the past 90 days. VIEW TWEETS
0 Extracted Citations
22 Extracted References
Similar Papers

Referenced Papers

Publications referenced by this paper.
Showing 1-10 of 22 references

The physical aspects of total and half body photon irradiation

  • J van Dyk, JM Galvin, GP Glasgrow, EB. Podgorsak
  • Task Group
  • 1986
Highly Influential
3 Excerpts

Monte Carlo treatment planning for a new Co-60 total body irradiator

  • M Snyder, R Halford, B Loughery, A Nalichowski, T Bossenberger, J. Burmeister
  • Med Phys. 2017;44:2918
  • 2017
2 Excerpts

Similar Papers

Loading similar papers…