Clinical digital breast tomosynthesis system: dosimetric characterization.

@article{Feng2012ClinicalDB,
  title={Clinical digital breast tomosynthesis system: dosimetric characterization.},
  author={Steve S. J. Feng and Ioannis Sechopoulos},
  journal={Radiology},
  year={2012},
  volume={263 1},
  pages={35-42}
}
PURPOSE To comprehensively characterize the dosimetric properties of a clinical digital breast tomosynthesis (DBT) system for the acquisition of mammographic and tomosynthesis images. MATERIALS AND METHODS Compressible water-oil mixture phantoms were created and imaged by using the automatic exposure control (AEC) of the Selenia Dimensions system (Hologic, Bedford, Mass) in both DBT and full-field digital mammography (FFDM) mode. Empirical measurements of the x-ray tube output were performed… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 41 extracted citations

Cancer risk estimation in Digital Breast Tomosynthesis using GEANT4 Monte Carlo simulations and voxel phantoms.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics • 2016
View 4 Excerpts
Highly Influenced

Optimal photon energy comparison between digital breast tomosynthesis and mammography: a case study.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics • 2014
View 4 Excerpts
Highly Influenced

Update on new technologies in digital mammography

International journal of women's health • 2014
View 9 Excerpts
Highly Influenced

How many physicists does it take to test a mammography unit?

Australasian Physical & Engineering Sciences in Medicine • 2017
View 1 Excerpt

References

Publications referenced by this paper.
Showing 1-10 of 26 references

A new generation FFDM / tomosynthesis fusion system with selenium detector

M Saito
Proc SPIE • 2010

Computation of the glandular radiation dose in digital tomosyn - thesis of the breast

RH Moore, DB Kopans, +3 authors ED Yeh
Med Phys • 2007