Characterization of the interface in the plasma-sprayed HA coating/Ti-6Al-4V implant system.

@article{Filiaggi1991CharacterizationOT,
  title={Characterization of the interface in the plasma-sprayed HA coating/Ti-6Al-4V implant system.},
  author={M J Filiaggi and Neil A Coombs and Robert M. Pilliar},
  journal={Journal of biomedical materials research},
  year={1991},
  volume={25 10},
  pages={1211-29}
}
The successful use of plasma-sprayed hydroxyapatite (HA) coatings on Ti-alloy implants for implant-to-bone fixation requires strong adherence of the ceramic coating to the underlying metal substrate. In this study, the metal-ceramic interface was evaluated using mechanical, chemical, and structural characterization methods. Evaluations of an HA-coated Ti-6Al-4V implant system using a modified short bar technique for interfacial fracture toughness determination revealed relatively low fracture… CONTINUE READING