Carboxyhaemoglobinaemia and pulse oximetry.

Abstract

We compared measurements obtained with a pulse oximeter (SpO2) against values obtained from a CO-oximeter in a patient with carbon monoxide poisoning. SpO2 was equal to the sum of the oxyhaemoglobin (HbO) and carboxyhaemoglobin (HbCO) values over the range of HbCO from 30 to 1%. This confirms the experimental findings that pulse oximeters measure HbCO as HbO. The patient was treated with oxygen (FlO2 = 50%) and recovered without any sequelae. Under these circumstances, the half-life of HbCO was approximately 2 h.

Statistics

02040'94'97'00'03'06'09'12'15
Citations per Year

100 Citations

Semantic Scholar estimates that this publication has 100 citations based on the available data.

See our FAQ for additional information.

Cite this paper

@article{Vegfors1991CarboxyhaemoglobinaemiaAP, title={Carboxyhaemoglobinaemia and pulse oximetry.}, author={M. Vegfors and Claes Lennmarken}, journal={British journal of anaesthesia}, year={1991}, volume={66 5}, pages={625-6} }