Bow-tie wobble artifact: effect of source assembly motion on cone-beam CT.

@article{Zheng2011BowtieWA,
  title={Bow-tie wobble artifact: effect of source assembly motion on cone-beam CT.},
  author={Dandan Zheng and John C Ford and Shun Wang and Dimitrios Lazos and Geoffrey D. Hugo and Dipendra Pokhrel and Lisha Zhang and Jeffrey F. Williamson},
  journal={Medical physics},
  year={2011},
  volume={38 5},
  pages={
          2508-14
        }
}
PURPOSE To investigate the cause of a bow-tie wobble artifact (BWA) discovered on Varian OBI CBCT images and to develop practical correction strategies. METHOD AND MATERIALS The dependence of the BWA on phantom geometry, phantom position, specific system, and reconstruction algorithm was investigated. Simulations were conducted to study the dependence of the BWA on scatter and beam hardening corrections. Geometric calibration was performed to rule out other gantry-angle dependent mechanical… CONTINUE READING