Biomechanical pull-out strength of the cervical spine locking plate (CSLP) expansion screws

@article{Mitchell1997BiomechanicalPS,
  title={Biomechanical pull-out strength of the cervical spine locking plate (CSLP) expansion screws},
  author={David W. Mitchell and Russel A Betcher and Stuart Littlejohn and L. Duarte Dillon and Robert A. Mcguire and Lyle D. Zardiackas},
  journal={Proceedings of the 1997 16 Southern Biomedical Engineering Conference},
  year={1997},
  pages={151-154}
}
This study was designed to evaluate the biomechanical pull-out strength and mechanism of pull-out in cantilever bending of the Synthes cervical spine locking plate (CSLP) using a porcine spine model of four different length 4.35 mm diameter expansion screws and the 3.5 mm bicortical cancellous screw with and without end plates removed. Ten constructs using each screw length were tested in single cycle cantilever bending by loading the free end of the plate in both flexion and extension, with… CONTINUE READING