Biomechanical evaluation of the dynamic hip screw with two- and four-hole side plates.

@article{Mcloughlin2000BiomechanicalEO,
  title={Biomechanical evaluation of the dynamic hip screw with two- and four-hole side plates.},
  author={S W Mcloughlin and Donna L. Wheeler and J Rider and Brett R Bolhofner},
  journal={Journal of orthopaedic trauma},
  year={2000},
  volume={14 5},
  pages={318-23}
}
OBJECTIVES To determine the biomechanical strength and stiffness of a dynamic hip screw (DHS; Synthes USA, Paoli, PA, U.S.A.) with a two-hole side-plate as compared with a four-hole side-plate design for the reconstruction of unstable three-part intertrochanteric fractures. DESIGN Eight matched pairs of embalmed human femurs were tested in two modes: (a) 2,000 cycles of simulated physiologic loading; (b) test to failure. SETTING Laboratory. Simulated single leg stance using a simulated… CONTINUE READING