Average glandular dose in paired digital mammography and digital breast tomosynthesis acquisitions in a population based screening program: effects of measuring breast density, air kerma and beam quality.

@article{sters2018AverageGD,
  title={Average glandular dose in paired digital mammography and digital breast tomosynthesis acquisitions in a population based screening program: effects of measuring breast density, air kerma and beam quality.},
  author={B. H. {\O}ster{\aa}s and P. Skaane and R. Gullien and A. C. Martinsen},
  journal={Physics in medicine and biology},
  year={2018},
  volume={63 3},
  pages={
          035006
        }
}
The main purpose was to compare average glandular dose (AGD) for same-compression digital mammography (DM) and digital breast tomosynthesis (DBT) acquisitions in a population based screening program, with and without breast density stratification, as determined by automatically calculated breast density (Quantra™). Secondary, to compare AGD estimates based on measured breast density, air kerma and half value layer (HVL) to DICOM metadata based estimates. AGD was estimated for 3819 women… Expand
Comparing Screening Outcomes for Digital Breast Tomosynthesis and Digital Mammography by Automated Breast Density in a Randomized Controlled Trial: Results from the To-Be Trial.
TLDR
Screening with digital breast tomosynthesis combined with synthetic two-dimensional mammograms (DBT+SM) versus digital mammography (DM) yielded lower recall rates for women with Volpara Density Grade (VDG) 1 and VDG 2, and adjusted relative risk of recall and screen-detected breast cancer increased with denser breasts for DBT+SM but not for DM. Expand
Digital Mammography versus Breast Tomosynthesis: Impact of Breast Density on Diagnostic Performance in Population-based Screening.
TLDR
Digital breast tomosynthesis enabled the detection of more cancers in all density and age groups compared with digital mammography, especially cancers classified as spiculated masses and architectural distortions. Expand
INSTITUTIONAL BREAST DOSES IN DIGITAL MAMMOGRAPHY.
TLDR
The estimated values of breast dose in screening mammography, establish institutional doses and compare them with the corresponding dose values were found to be comparable, or in most cases lower, than the corresponding 75th and 95th percentile values from previous studies. Expand
Additive value of exposure parameters for breast cancer diagnosis in digital mammography
TLDR
Investigating the effect of different breast lesions on exposure parameters in digital mammography and whether the exposure parameters can additively improve diagnostic efficiency found exposure parameters additively improved diagnostic accuracy for breast cancer. Expand
Accuracy of mammography dosimetry in the era of the European Directive 2013/59/Euratom transposition.
TLDR
Pat-specific mean glandular dose allows more accurate individual breast dosimetry than what would be performed using organ dose provided by manufacturers, and definition of diagnostic reference levels could be achieved using either psMGD or organ dose. Expand
A randomized controlled trial of digital breast tomosynthesis versus digital mammography in population-based screening in Bergen: interim analysis of performance indicators from the To-Be trial
TLDR
In this RCT, DBT was associated with longer interpretation time than DM and recall rates were lower for DBT than for DM, which showed that DBT took longer to read than DM, but had significantly lower recall rate than DM. Expand
Diagnostic Performance of Digital Breast Tomosynthesis for Breast Suspicious Calcifications From Various Populations: A Comparison With Full-field Digital Mammography
TLDR
In this study, DBT exhibited a superior advantage in dense breasts and benign calcification cases compared to FFDM, while no advantage was observed in non-dense breasts or malignant calcifications cases. Expand
Association of the Differences in Average Glandular Dose with Breast Cancer Risk
  • Le Ma, Yuxing Cai, +4 authors G. Qin
  • Medicine
  • 2020
TLDR
NAGD differences between both breasts can be included in the breast cancer risk assessment model to evaluate breast cancer chance and improves breast cancer classification. Expand
Digital Breast Tomosynthesis: Concepts and Clinical Practice.
TLDR
The authors review the benefits of DBT imaging in screening and diagnostic breast imaging and the quasi three-dimensional information obtained from the reconstructed DBT data set allows a more efficient imaging work-up than imaging with two-dimensional full-field digital mammography alone. Expand
Modèle biomécanique du sein pour l’évaluation de la compression et de la perception d’inconfort en mammographie
Contexte: La mammographie est une modalite d'imagerie medicale a faible dose permettant la detection du cancer mammaire a un stade precoce. Lors de l'examen, le sein est comprime entre deux plaquesExpand

References

SHOWING 1-10 OF 32 REFERENCES
Automated Breast Density Computation in Digital Mammography and Digital Breast Tomosynthesis: Influence on Mean Glandular Dose and BIRADS Density Categorization.
TLDR
Breast density measurements from both algorithms lead to equivalent BIRADS classification and MGD values, hence showing no difference in clinical outcomes. Expand
Estimation of percentage breast tissue density: comparison between digital mammography (2D full field digital mammography) and digital breast tomosynthesis according to different BI-RADS categories.
TLDR
Using DBT, breast density values were lower than those obtained using 2D FFDM, with a non-linear relationship across the classical BI-RADS categories, which are relevant for clinical practice and research studies using density in determining the risk. Expand
Average glandular dose in digital mammography and breast tomosynthesis.
  • T. Olgar, T. Kahn, D. Gosch
  • Medicine
  • RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin
  • 2012
TLDR
The mean calculated AGD per exposure in 3 D imaging mode was on average 34% higher than for 2 S imaging mode for patients examined with the same CBT, and a good correlation coefficient was found. Expand
Radiation dose with digital breast tomosynthesis compared to digital mammography: per-view analysis
TLDR
These findings show a modest increase of radiation dose to the breast by tomosynthesis compared to FFDM, and its use in conjunction with synthetic 2D images should not be deterred by concerns regarding radiation burden, and should draw on evidence of potential clinical benefit. Expand
Review of radiation dose estimates in digital breast tomosynthesis relative to those in two-view full-field digital mammography.
TLDR
How radiation dose levels in digital breast tomosynthesis (DBT) differ from those used in 2-view full-field digital mammography (FFDM) is examined, with substantial implications for population screening programs. Expand
Comparative evaluation of average glandular dose and breast cancer detection between single-view digital breast tomosynthesis (DBT) plus single-view digital mammography (DM) and two-view DM: correlation with breast thickness and density
TLDR
DBT has higher diagnostic performance and potential to overcome limitations of DM, and MLO DBT plus CC DM provided higher diagnosticperformance than two-view DM in dense breasts with a small increase in AGD. Expand
'In vivo' average glandular dose evaluation: one-to-one comparison between digital breast tomosynthesis and full-field digital mammography.
TLDR
Examination parameter data collection and 'in vivo' dosimetry using metal oxide semiconductor field effect transistor (MOSFET) were utilised to determine and compare entrance skin air kerma (ESAK) and average glandular dose (AGD) from a radioprotection viewpoint in the DBT and COMBO modalities. Expand
Average glandular dose in digital mammography and digital breast tomosynthesis: comparison of phantom and patient data.
TLDR
It was found that the ratios between patient and phantom AGD did not differ considerably using both dosimetry phantoms, and these ratios were deemed to be sufficiently close to unity to be suitable for Dosimetry evaluation in quality control procedures. Expand
Two-view digital breast tomosynthesis screening with synthetically reconstructed projection images: comparison with digital breast tomosynthesis with full-field digital mammographic images.
TLDR
The combination of current reconstructed 2D images and DBT performed comparably to FFDM plus DBT and is adequate for routine clinical use when interpreting screening mammograms. Expand
Patient Specific Dose Calculation Using Volumetric Breast Density for Mammography and Tomosynthesis
TLDR
A method to calculate the patient-specific MGD is described using a representation of the patient’s volumetric breast density to derive the breast glandularity, which personalises the MGD to the individual woman, rather than assuming a constant value, or one that depends solely on compressed breast thickness. Expand
...
1
2
3
4
...