Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy: an evaluation of the setup verification process for the special case of a rigid head phantom.

@article{Boswell2006AutomaticRO,
  title={Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy: an evaluation of the setup verification process for the special case of a rigid head phantom.},
  author={Sarah A Boswell and Wolfgang A. Tom{\'e} and Robert Jeraj and Hazim A Jaradat and Thomas Rock Mackie},
  journal={Medical physics},
  year={2006},
  volume={33 11},
  pages={4395-404}
}
Precise daily target localization is necessary to achieve highly conformal radiation delivery. In helical tomotherapy, setup verification may be accomplished just prior to delivering each fraction by acquiring a megavoltage CT scan of the patient in the treatment position. This daily image set may be manually or automatically registered to the image set on which the treatment plan was calculated, in order to determine any needed adjustments. The system was tested by acquiring 104 MVCT scans of… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 28 extracted citations

Similar Papers