Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years.

@article{Kleemann2007AnnualRO,
  title={Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years.},
  author={Thomas Kleemann and Torsten C. Becker and Klaus Doenges and Margit Vater and Jochen Senges and Steffen Schneider and Werner Saggau and Udo Weisse and Karlheinz Seidl},
  journal={Circulation},
  year={2007},
  volume={115 19},
  pages={2474-80}
}
BACKGROUND The number of patients with longer follow-up after implantation of an implantable cardioverter-defibrillator is increasing continuously. Defibrillation lead failure is a typical long-term complication. Therefore, the long-term reliability of implantable cardioverter-defibrillator leads has become an increasing concern. The aim of the present study was to assess the annual rate of transvenous defibrillation lead defects related to follow-up time after lead implantation. METHODS AND… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 235 extracted citations

15 Complications of Pacemaker Implantation

Michele Schiariti, Mariateresa Cacciola, Paolo Emilio Puddu
2017
View 5 Excerpts
Highly Influenced

Complications of Pacemaker Implantation

Mariateresa Cacciola
2012
View 5 Excerpts
Highly Influenced

Is the use of an additional pace/sense lead the optimal strategy for the avoidance of lead extraction in defibrillation lead failure? A single-centre experience.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology • 2010
View 4 Excerpts
Highly Influenced

Potential role of home monitoring to reduce inappropriate shocks in implantable cardioverter-defibrillator patients due to lead failure.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology • 2009
View 4 Excerpts
Highly Influenced

Cardiac rhythm management devices

Australian journal of general practice • 2018

References

Publications referenced by this paper.
Showing 1-10 of 30 references

Long-term follow-up of transvenous defibrillation leads: high incidence of fracture in coaxial polyurethane lead

S Kitamura, S Kazuhiro, +5 authors K. Shiro
Circ J. 2006;70:273–277 • 2006

Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction

AE Buxton, KL Lee, +9 authors G Hafley
N Engl J Med • 2002