Analytical evaluation of the clinical chemistry analyzer Olympus AU 2700 plus

@inproceedings{Juricek2010AnalyticalEO,
  title={Analytical evaluation of the clinical chemistry analyzer Olympus AU 2700 plus},
  author={Jasna Juricek and Lovorka Derek and Adriana Unic and Tihana Serdar and Domagoj Marijan{\vc}evi{\'c} and Marcela Zivkovi{\'c} and Zeljko Romi{\'c}},
  year={2010}
}
Materials and methods: The evaluation consisted of determination of within-run and between-run imprecision, inaccuracy and comparison with Olympus AU2700. The tested analytes were: glucose, creatinine, urate, total bilirubin, cholesterol, tryglicerides, calcium, phosphate, iron, unsaturated iron binding capacity, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, creatine kinase, alpha-amylase, alkaline phosphatase, potassium, sodium, chloride, C-reactive protein… CONTINUE READING