• Medicine
  • Published in Folia medica 1997

Analytic performance of digoxin laboratory monitoring.

@article{Dimitrova1997AnalyticPO,
  title={Analytic performance of digoxin laboratory monitoring.},
  author={Roumiana Dimitrova and Nasko Atanassov},
  journal={Folia medica},
  year={1997},
  volume={39 2},
  pages={
          10-4
        }
}
Since therapeutic drug monitoring is rapidly becoming a widely-used tool in clinical medicine we prospectively assessed 547 control samples over a period of 18 months using fluorescence polarization immunoassay, calibrators and control samples to determine the analytical performance of Abbott TDx in measuring serum digoxin levels. Recovery of the controls (within-run and between-run coefficient of variation) evaluated at the low, target therapeutic and intoxication ranges was 14%, 8% and 7% and… CONTINUE READING