Analysis of beta-hydroxy-beta-methyl butyrate in plasma by gas chromatography and mass spectrometry.

@article{Nissen1990AnalysisOB,
  title={Analysis of beta-hydroxy-beta-methyl butyrate in plasma by gas chromatography and mass spectrometry.},
  author={Steven L. Nissen and Mark Van Koevering and D Webb},
  journal={Analytical biochemistry},
  year={1990},
  volume={188 1},
  pages={
          17-9
        }
}
A method for measuring the branched chain hydroxy acid beta-hydroxy-beta-methyl butyrate (HMB, a product of leucine catabolism) has been described. A [2H6]HMB internal standard was added to plasma and standards, and samples were extracted with diethyl ether, backextracted into neutral phosphate, dried, and derivatized for gas chromatography and mass spectrometry. The natural HMB was monitored at 175 amu and the deuterated HMB was monitored at 181 amu. Standard curves were linear to at least 25… CONTINUE READING