An automated biotin-streptavidin procedure for progesterone evaluation.

Abstract

An automated biotin-streptavidin procedure for measuring progesterone in serum is described. The method is linear up to 98.3 nmol/L and the calibration curve is stable for at least 14 days. The lower limit of progesterone detection is 0.70 nmol/L. Although serum is the preferred specimen, progesterone levels can be measured in EDTA, heparin, or citrated… (More)