Accurate transfer function determination for superconducting gravimeters

@article{Camp2000AccurateTF,
  title={Accurate transfer function determination for superconducting gravimeters},
  author={M. Camp and H. Wenzel and P. Schott and P. Vauterin and O. Francis},
  journal={Geophysical Research Letters},
  year={2000},
  volume={27},
  pages={37-40}
}
The transfer function for the cryogenic gravimeter GWR-C021 operating at Membach (Belgium) has been experimentally determined by injecting known voltages into the control electronics of the system. The output of the gravimeter to the injected sine waves and step functions has been observed. This give a precise knowledge of the transfer function of the gravimeter. It allows one to reach a precision of better than 0.01 second in the phase response of the instrument, in agreement with the Global… Expand
51 Citations
Comparison between the Transfer Functions of three Superconducting Gravimeters
  • 5
  • PDF
3.04 – Superconducting Gravimetry
  • 24
Optimized strategy for the calibration of superconducting gravimeters at the one per mille level
  • 22
  • PDF
On the Accuracy of the Calibration of Superconducting Gravimeters Using Absolute and Spring Sensors: a Critical Comparison
  • 19
...
1
2
3
4
5
...