Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor.

@article{Baumgart1998AccuracyOT,
  title={Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor.},
  author={Baumgart and Kamp},
  journal={Blood pressure monitoring},
  year={1998},
  volume={3 5},
  pages={303-307}
}
OBJECTIVE: To test the SpaceLabs Medical 90217 ambulatory blood pressure monitor for compliance with The Association for the Advancement of Medical Instrumentation's standard and a modification of the British Hypertension Society (BHS) protocol. METHODS: The study concerned 85 subjects aged 17-88 years with various arm circumferences and ranges of blood pressure. Resting casual blood pressure was 135 +/- 24/76 +/- 14 mmHg (mean +/- SD). For each subject nine readings of the 90217 device were… CONTINUE READING