A systematic review of variability and reliability of manual and automated blood pressure readings.

@article{Skirton2011ASR,
  title={A systematic review of variability and reliability of manual and automated blood pressure readings.},
  author={Heather Skirton and Wendy Chamberlain and C Lawson and Helen Ryan and Emma Young},
  journal={Journal of clinical nursing},
  year={2011},
  volume={20 5-6},
  pages={
          602-14
        }
}
AIMS AND OBJECTIVES To compare the accuracy and appropriateness of auscultatory (manual) and oscillometric (automated) devices for measuring blood pressure in clinical settings. BACKGROUND Accurate measurement of blood pressure is integral to early recognition of deterioration in the condition of a patient. Despite recommendations regarding the use of auscultatory devices in situations where treatment decisions are made dependent on blood readings, the use of automated machines is becoming… 
INACCURACY OF NONINVASIVE BLOOD PRESSURE MONITOR DEVICES, DO WE HAVE AN ALTERNATIVE?
TLDR
It is suggested that automated method in measurement of BP frequently shows lower BP, which can influence diagnosis of hypertension and management of certain medical conditions.
Sources of inaccuracy in the measurement of adult patients’ resting blood pressure in clinical settings: a systematic review
TLDR
A single BP value outside the expected range should be interpreted with caution and not taken as a definitive indicator of clinical deterioration, where a measurement is abnormally high or low, further measurements should be taken and averaged.
Accuracy Assessment of Measured Blood Pressures in Three Departments of one Tertiary Hospital
TLDR
One of the possible causes of difference between routine and standard measured blood pressures in the present study was lack of calibration in a sphygmomanometer and this defect was also the easier problem to correct by change of dam.
Impact on variation in the direction of cuff bladder on blood pressure readings
TLDR
The test procedure showed agreement with the standard protocol of cuff positioning, therefore, it could be used interchangeably and could beused interchangeably for BP measurement.
(In)accuracy of blood pressure measurement in 14 Italian hospitals
TLDR
The operator's compliance to some recommendations in BP measurement is unacceptably low, and this survey provides detailed indications for medical directors on the procedures and settings to prioritize in educational programs, which are definitely needed.
The inaccuracy of automatic devices taking postural measurements in the emergency department.
TLDR
It is suggested that automatic devices cannot reliably detect or rule out orthostatic hypotension, indicating that triage nurses need to use manual devices to take accurate postural blood pressures for optimal patient care.
Automated Versus Manual Blood Pressure Measurement: A Randomized Crossover Trial in the Emergency Department of a Tertiary Care Hospital in Karachi, Pakistan: Are Third World Countries Ready for the Change?
TLDR
Systolic readings from a previously validated device are not reliable when used in the ER and they show a higher degree of incongruency and inaccuracy when they are used outside validation settings.
Effects of room environment and nursing experience on clinical blood pressure measurement: an observational study
TLDR
The results indicate that frequent nursing trainings are important for nurses to achieve accurate manual auscultatory blood pressure measurement.
Reproducibility of visit-to-visit variability of blood pressure measured as part of routine clinical care
TLDR
Primary analysis of clinical trial data suggests visit-to-visit variability (VVV) of blood pressure is strongly associated with the incidence of cardiovascular disease and VVV of SBP measured in a real-world setting is not random.
...
1
2
3
4
5
...

References

SHOWING 1-10 OF 39 REFERENCES
Clinical evaluation of an automated oscillometric blood pressure wrist device
TLDR
The Omron Rx wrist monitor is accurate according to the British Hypertension Society protocol, but narrowly fails the Association of the Advancement of Medical Instrumentation criteria.
Hospital blood pressure measurement: staff and device assessment.
Evaluation of the ability of clinical staff to measure blood pressure as well as the functional state of hospital sphygmomanometers has consistently demonstrated marked deficiencies. In this study,
Accuracy of an automated blood pressure device in stable inpatients: optimum vs routine use.
TLDR
The IVAC 4200 yields substandard estimates of systolic and diastolic BP even under standardized, thus optimum conditions and the presence of arrhythmias or low K5 values and the selection of inappropriate cuff size by the ward staff also contributed to inaccuracy.
In-use comparison of blood pressure measurements from an automated blood pressure instrument with those from a mercury sphygmomanometer.
TLDR
This study evaluated the ABPMI used in a large, urban, tertiary-care teaching hospital in the northeastern United States under routine conditions and suggested a strong association between the methods and Bland-Altman analysis indicated an acceptably small systematic error but greater random error.
Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and the wrist.
TLDR
Average sphygmomanometer values exceed oscillometrically measured blood pressure values but individual disagreements cannot be predicted.
Automated assessment of blood pressure using BpTRU compared with assessments by a trained technician and a clinic nurse
TLDR
The BpTRU had desirable characteristics that suggest that it would be a suitable replacement for auscultatory assessment of blood pressure in clinical practice and a large confirmatory study performed in a usual clinic setting is required.
Are automated blood pressure measurements accurate in trauma patients?
TLDR
Injury severity, degree of acidosis, and resuscitation volume were more accurately reflected by manual BP, particularly in hypotensive patients, and automated BP determinations were consistently higher than manual BP.
Validation of the DINAMAP ProCare blood pressure device according to the international protocol in an adult population
TLDR
The DINAMAP ProCare Monitor can be recommended for clinical use in an adult population and falls within the passing criteria for the Association for the Advancement of Medical Instrumentation Standard (ANSI/AAMI SP-10).
The validity of emergency department triage blood pressure measurements.
TLDR
ABP triage measurements show significant discrepancies from a reference standard, and repeat measurements following AHA standards demonstrate significant decreases in the measured blood pressures.
Evaluation of a community-based automated blood pressure measuring device.
TLDR
Neither automated device met the AAMI or BHS criteria for accuracy while in use in the community, and neither performed as well in thecommunity as in the laboratory.
...