A rapid reversed phase high performance liquid chromatographic method for determination of etoposide (VP-16) in human plasma.

@article{Shirazi2001ARR,
  title={A rapid reversed phase high performance liquid chromatographic method for determination of etoposide (VP-16) in human plasma.},
  author={Farshad Hoseini Shirazi and Gholamreza Golmohammadnajad Bahrami and David John Stewart and Eva Tomiak and Franqois Delorme and Denis Noel and Rakesh Goel},
  journal={Journal of pharmaceutical and biomedical analysis},
  year={2001},
  volume={25 3-4},
  pages={
          353-6
        }
}
A rapid, simple and sensitive isocratic High Performance Liquid Chromatography (HPLC) method was developed to measure the concentration of etoposide in plasma samples with UV detection at 220 nm. The method uses a Bondapac C18 column at 60 degrees C. The mobile phase consists of Methanol: water (45:55 v/v) at a flow rate of 2.8 ml/min. Phenacetin was used as an internal standard. The plasma samples were extracted using ether with the organic layer evaporated under nitrogen. The residue was… CONTINUE READING