A new method for quantitative ultrasound measurements at multiple skeletal sites: first results of precision and fracture discrimination.

@article{Barkmann2000ANM,
  title={A new method for quantitative ultrasound measurements at multiple skeletal sites: first results of precision and fracture discrimination.},
  author={Reinhard Barkmann and Efraim Olszewer Kantorovich and C Singal and Didier B. Hans and Harry K. Genant and Markus Heller and Claus C. Gl{\"u}er},
  journal={Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry},
  year={2000},
  volume={3 1},
  pages={1-7}
}
We investigated a new multisite quantitative ultrasound device that measures the acoustic velocity in axial transmission mode along the cortex. Using a prototype of the Omnisense (Sunlight Ultrasound Technologies, Rehovot, Israel), we tested the performance of this instrument at four sites of the skeleton: radius, ulna, metacarpal, and phalanx. Intraobserver (interobserver) precision errors ranged from 0.2% to 0.3% (0.3% to 0.7%) for triplicate measurements with repositioning. Fracture… CONTINUE READING