A low noise thermometer readout for ruthenium oxide resistors

@inproceedings{Fixsen2002ALN,
  title={A low noise thermometer readout for ruthenium oxide resistors},
  author={D. J. Fixsen and P. Mirel and Alan Kogut and Michael Dean Seiffert},
  year={2002}
}
  • D. J. Fixsen, P. Mirel, +1 author Michael Dean Seiffert
  • Published 2002
  • Physics
  • The thermometer readout and thermal control system for the Absolute Radiometer for Cosmology, Astrophysics, and Diffuse Emission (ARCADE) experiment is described, including the design, testing, and results from the first flight of ARCADE. The noise is equivalent to about 1 Ω or 0.15 mK in a second for the RuO2 resistive thermometers at 2.7 K. The average power dissipation in each thermometer is 1 nW. The control system can take full advantage of the thermometers to maintain stable temperatures… CONTINUE READING

    Citations

    Publications citing this paper.

    References

    Publications referenced by this paper.
    SHOWING 1-2 OF 2 REFERENCES

    Cryogenics

    • C Baciocco, Boragno, Valbusa
    • Cryogenics
    • 1989

    MR Corson, Rev. Sci. Inst

    • MR Corson, Rev. Sci. Inst
    • 1985