A fast 3D look-locker method for volumetric T1 mapping.

@article{Henderson1999AF3,
  title={A fast 3D look-locker method for volumetric T1 mapping.},
  author={Eduardo Henderson and Graeme McKinnon and Thomas Youbok Lee and Brian K. Rutt},
  journal={Magnetic resonance imaging},
  year={1999},
  volume={17 8},
  pages={
          1163-71
        }
}
We introduce a fast technique, based on the principles of the 2D Look-Locker T1 measurement scheme, to rapidly acquire the data for accurate maps of T1 in three dimensions. The acquisition time has been shortened considerably by segmenting the acquisition of the k(y) phase encode lines. Using this technique, the data for a 256 x 128 x 32 volumetric T1 measurement can be acquired in 7.6 min. T1 measurements made in phantoms with T1s between 200 and 1200 ms had an accuracy of 4% and a… CONTINUE READING

Topics from this paper.

Citations

Publications citing this paper.
SHOWING 1-10 OF 64 CITATIONS, ESTIMATED 63% COVERAGE

FILTER CITATIONS BY YEAR

2003
2019

CITATION STATISTICS

  • 2 Highly Influenced Citations