A cone-beam tomography algorithm for orthogonal circle-and-line orbit.

@article{Zeng1992ACT,
  title={A cone-beam tomography algorithm for orthogonal circle-and-line orbit.},
  author={G L Zeng and Grant T. Gullberg},
  journal={Physics in medicine and biology},
  year={1992},
  volume={37 3},
  pages={563-77}
}
A cone-beam algorithm which provides a practical implementation of B D Smith's cone-beam inversion formula is presented. For a cone-beam vertex orbit consisting of a circle and an orthogonal line. This geometry is easy to implement in a SPECT system, and it satisfies the cone-beam data sufficiency condition. The proposed algorithm is in the form of a convolution-back projection, and requires a pre-filtering procedure. Computer simulations show a reduction of the artifacts that are found with… CONTINUE READING

From This Paper

Topics from this paper.
30 Citations
13 References
Similar Papers

Citations

Publications citing this paper.
Showing 1-10 of 30 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 13 references

An inversion formula far cone-beam remnstwtion

  • H K
  • SIAM J. Appl. Math
  • 1983
Highly Influential
3 Excerpts

Derivation and analysis of a filtered backprojection algorithm for cone beam reconstruction methods IEEE Trans

  • H YanX, M LeahyR
  • 1991

SPECT using a specially desiened cone beam Collimator 3

  • K JaszczakRJ.Greer, E ColemanR
  • Nucl . Med
  • 1988

Similar Papers

Loading similar papers…