Glass-electrode assemblies in which the reference half-cell contains a porous ceramic type of liquid junction are likely to produce misleading pH measurements under normal service conditions. The error arises from substantial liquid-junction potentials, associated with the porous ceramic plug, which vary with the ionic composition of the solution under test. The error is not revealed by conventional two-point calibration procedures, since the majority of standard buffer solutions have a similar… CONTINUE READING