A Comparison of Ansoft HFSS and CST Microwave Studio Simulation Software for Multi-channel Coil Design and SAR Estimation at 7 T MRI

@inproceedings{Kozlov2010ACO,
  title={A Comparison of Ansoft HFSS and CST Microwave Studio Simulation Software for Multi-channel Coil Design and SAR Estimation at 7 T MRI},
  author={Mikhail Kozlov and Robert Turner},
  year={2010}
}
We present a comparison of RF simulations with Ansoft HFSS and CST Microwave studio software, for a commercially available 7 T multi-channel coil. We investigated the conditions influencing precision and simulation time. When simulation parameters (such as mesh size) are appropriately selected and full coverage of the results is ensured, there is agreement between simulated and experimentally determined S parameters, between simulated and actual capacitor values, and between simulated and… CONTINUE READING
Highly Cited
This paper has 46 citations. REVIEW CITATIONS
13 Citations
2 References
Similar Papers

Citations

Publications citing this paper.
Showing 1-10 of 13 extracted citations

References

Publications referenced by this paper.
Showing 1-2 of 2 references

Optimization of SAR calculation for 3-D EM time and frequency domain data,

  • M. Kozlov, R. Turner
  • Proceedings of International Society for Magnetic…
  • 2009
1 Excerpt

Similar Papers