• Corpus ID: 75904609

A Comparative Study on the Results of Liver Function between Olympus AU1000 and Backman CX7 Detection Systems

@article{Bo2006ACS,
  title={A Comparative Study on the Results of Liver Function between Olympus AU1000 and Backman CX7 Detection Systems},
  author={Wang Bo},
  journal={Chinese Journal of Evidence-Based Medicine},
  year={2006}
}
  • W. Bo
  • Published 2006
  • Medicine
  • Chinese Journal of Evidence-Based Medicine
Objective To study the comparability of liver function results between the two detection systems. Methods Based on the NCCLS EP9-A document, 8 fresh serum samples were collected daily for the assay of 10 routine liver function parameters by utilizing the Olympus AU1000 and Backman CX7 detection systems respectively. The results were recorded over 5 days consecutively. Linearity equation and relative deviations were calculated. The comparability between the results obtained from different…