6D image guidance for spinal non-invasive stereotactic body radiation therapy: Comparison between ExacTrac X-ray 6D with kilo-voltage cone-beam CT.

@article{Chang20106DIG,
  title={6D image guidance for spinal non-invasive stereotactic body radiation therapy: Comparison between ExacTrac X-ray 6D with kilo-voltage cone-beam CT.},
  author={Zheng Chang and Zhiheng Wang and Jinli Ma and Jennifer C. O'Daniel and John P. Kirkpatrick and Fang-Fang Yin},
  journal={Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology},
  year={2010},
  volume={95 1},
  pages={116-21}
}
PURPOSE To investigate setup discrepancies measured with ExacTrac X-ray 6 degree-of-freedom (6D) and cone-beam computed tomography (CBCT) for patients under treatments of stereotactic body radiation therapy (SBRT). MATERIALS AND METHODS In this work, phantom and patient studies were performed. In the phantom studies, an anthropomorphic phantom was placed with pre-defined positions, and imaged with ExacTrac X-ray 6D and CBCT to test the accuracy of the imaging systems. In the patient studies… CONTINUE READING