2.5-km Low-Temperature Multiple-Reflection Cell.

@article{Horn197125kmLM,
  title={2.5-km Low-Temperature Multiple-Reflection Cell.},
  author={D. Horn and G. Pimentel},
  journal={Applied optics},
  year={1971},
  volume={10 8},
  pages={
          1892-8
        }
}
  • D. Horn, G. Pimentel
  • Published 1971
  • Materials Science, Medicine
  • Applied optics
  • A low-temperature multiple-pass White cell is described which can be operated at any temperature between 120 K and room temperature. The path length can be varied from 60 m up to 2540 m in 80-m steps. The four-row multiple reflection system, a modified White arrangement, is enclosed in an aluminum dewar and can be focused externally while cold. An approximate expression for the astigmatic image distortion in a cell of the described type is derived. 