P. Ram Sukumar

Learn More
In this study, we investigated the correlation between 2-dimensional (2D) and 3D gamma analysis using the new PTW OCTAVIUS 4D system for various parameters. For this study, we selected 150 clinically approved volumetric-modulated arc therapy (VMAT) plans of head and neck (50), thoracic (esophagus) (50), and pelvic (cervix) (50) sites. Individual(More)
A commercial 2D array seven29 detector has been characterized and its performance has been evaluated. 2D array ionization chamber equipped with 729 ionization chambers uniformly arranged in a 27 × 27 matrix with an active area of 27 × 27 cm² was used for the study. An octagon-shaped phantom (Octavius Phantom) with a central cavity is used to insert the 2D(More)
The recalculation of 1 fraction from a patient treatment plan on a phantom and subsequent measurements have become the norms for measurement-based verification, which combines the quality assurance recommendations that deal with the treatment planning system and the beam delivery system. This type of evaluation has prompted attention to measurement(More)
AIM In this study, the dosimetric properties of the electronic portal imaging device were examined and the quality assurance testing of Volumetric Modulated Arc Therapy was performed. BACKGROUND RapidArc involves the variable dose rate, leaf speed and the gantry rotation. The imager was studied for the effects like dose, dose rate, field size, leaf speed(More)
AIM To compare and evaluate the performance of two different volumetric modulated arc therapy delivery techniques. BACKGROUND Volumetric modulated arc therapy is a novel technique that has recently been made available for clinical use. Planning and dosimetric comparison study was done for Elekta VMAT and Varian RapidArc for different treatment sites. (More)
Independent monitor unit verification calculation (MUVC) has been recommended by several authors for intensity modulated radiotherapy (IMRT) as a patient specific quality assurance tool. Aim of the present work is to develop an in-house excel spread sheet based MUVC program for volumetric modulated arc therapy (VMAT) using Clarkson's integration technique.(More)
This proposal proposes a DC/DC Buck Boost converter which has been used as a smooth starter for a DC Permanent Magnet Motor. In the existing system the DC/DC Buck Converter is used which provide the output less than the input Signal. Using buck converter it is difficult to increase the value of the input signal. Hence DC/DC BuckBoost Converter used from(More)
Kilo-voltage cone beam computed tomography (kV-CBCT) integrated with a linac can produce online volumetric and anatomical images for patient set-up and dosimetric analysis in adaptive radiotherapy. However CBCT is prone to motion artifacts. This study investigates the impact of target motion in CBCT imaging. To simulate respiratory movement, a dynamic(More)
AIM In measuring exit fluences, there are several sources of deviations which include the changes in the entrance fluence, changes in the detector response and patient orientation or geometry. The purpose of this work is to quantify these sources of errors. BACKGROUND The use of the volumetric modulated arc therapy treatment with the help of image(More)
Phase information recovered through interferometric techniques is mathematically wrapped in the interval (-π, π). Obtaining the original unwrapped phase is very important in numerous number of applications. This paper discusses a Fourier transform based phase unwrapping method. Kalman filter is proposed for denoising in post processing step to(More)