An electrostatic quadrupole lens has been substituted for a cylindrical lens system used in the original inductively coupled plasma-time-of-flight mass spectrometer (ICP-TOFMS). With an improved vacuum system also installed, the cylindrical and quadrupole lenses are compared to each other and to the performance of the prototype ICP-TOFMS. The quadrupole(More)
A time-of-flight mass spectrometer (TOFMS) was evaluated as a mass analyzer for inductively coupled plasma mass spectrometry (ICP-MS). The long-term drift of signals was in the range of 7-8% relative standard deviation, whereas the short-term precision was between 5 and 20%, somewhat worse than is typically reported for commercial ICP-MS instruments (5%).(More)
Isotope ratios and abundance sensitivities have been determined with an inductively coupled plasma-time-of-flight mass spectrometer (ICP-TOFMS). Abundance sensitivities are at least in the 10(6) range for low abundance ions that precede high abundance ions. Three methods of detection for isotope-ratio measurement have been compared. The three systems(More)
The coupling of an electrothermal vaporization (ETV) apparatus to an inductively coupled plasma time-of-flight mass spectrometer (ICP-TOFMS) is described. The ability of the ICP-TOFMS to produce complete elemental mass spectra at high repetition rates is experimentally demonstrated. A signal-averaging data acquisition board is employed to rapidly record(More)