"The low flow leak test" is recommended for pre-anesthetic inspection of anesthetic machines. We carried out anesthesia compression tests as a standard. Even in that case, often the low flow leak test does not meet the standard. We investigated the point where there is a leak in the anesthetic machine. Observing the leak that fluctuates each time there is(More)
To assess the current state of leak in anesthetic machines, we selected 66 units of anesthetic machines for inspection and repair from various medical institutions. Based on a newly designed inspection flow chart a low flow leak test for internal circuits of the anesthetic machines was performed. The conventional low flow leak test was also performed for(More)