PURPOSE In general, beam data of individual linac is measured for independent dose verification software program and the verification is performed as a secondary check. In this study, independent dose verification using golden beam data was compared to that using individual linac's beam data. METHODS Six institutions were participated and three different(More)
PURPOSE A multi-institutional study was performed to identify the impact of different independent dose verification programs on independent dose verification software program. METHODS Data for 1,543 treatment fields were collected in three institutions. RADCALC and Simple MU Analysis (Simple MU) using the Clarkson-based algorithm were used. RADCALC needs(More)
It is essential for quality assurance to verify the safety of each individual patient's plan in radiation therapy. The tolerance level for independent verification of monitor unit calculations for non-IMRT clinical radiotherapy has been shown in the AAPM TG114. Thus, we investigated the precision of independent MU (dose) verification considering a wedge(More)
PURPOSE CyberKnife irradiation is composed of tiny-size, multiple and intensity-modulated beams compared to conventional linacs. Few of the publications for Independent dose calculation verification for CyberKnife have been reported. In this study, we evaluated the feasibility of independent dose verification for CyberKnife treatment as Secondary check. (More)