The two-piece Van Straten Leuwen Poeschmann Metal (LPM) prosthesis was intended for the proximal interphalangeal joints. However, revision rates of 29% after 19 months were reported, as well as massive osteolysis. Five failed LPM titanium-niobium coated cobalt chromium components were obtained, three distal and two proximal, and subjected to a forensic(More)
Press-fit acetabular shells are subject to compressive loading conditions upon insertion during surgery. These conditions may result in deformation of the shell, potentially adversely affecting the performance of any liner fitted within it. This technical note describes the method used to measure acetabular shell deformation in a cadaver study using an(More)
Clinical results of the PyroCarbon proximal interphalangeal joint replacement are inconsistent with various complications reported. To address this, in vitro testing was conducted using finger joint simulators. Two PyroCarbon proximal interphalangeal prostheses were tested in a lubricant of dilute bovine serum to 5 × 10(6) cycles of flexion-extension(More)
Concerns have been raised that deformation of acetabular shells may disrupt the assembly process of modular prostheses. In this study we aimed to examine the effect that the strength of bone has on the amount of deformation of the acetabular shell. The hypothesis was that stronger bone would result in greater deformation. A total of 17 acetabular shells(More)
Deformation of the acetabular shell at the time of surgery can result in poor performance and early failure of the hip replacement. The study aim was to validate an ATOS III Triple Scan optical measurement system against a co-ordinate measuring machine using in vitro testing and to check repeatability under cadaver laboratory conditions. Two sizes of(More)
In contrast to the articulating and taper surfaces of failed total hip replacements, volumetric wear analysis of trunnions is not routinely performed. Metal wear particles from the trunnion may contribute not only to the failures of metal-on-metal total hip replacements but also to all hip replacements utilising metal trunnions. A validation study was(More)
BACKGROUND Large-diameter (≥36-mm) metal-on-metal (MoM) total hip replacements have been shown to fail at an unacceptably high rate. Globally, the DePuy Pinnacle prosthesis was the most widely used device of this type. There is evidence to suggest that one of the main reasons for the poor clinical performance of large-diameter MoM prostheses is the metal(More)
Recent guidance recommends the use of a well-proven cemented femoral stem for hemiarthroplasty in the management of fractures of the femoral neck, and the Exeter Trauma Stem (ETS) has been suggested as an example of such an implant. The design of this stem was based on the well-proven Exeter Total Hip Replacement stem (ETHRS). This study assessed the(More)
Press-fit acetabular shells used for hip replacement rely upon an interference fit with the bone to provide initial stability. This process may result in deformation of the shell. This study aimed to model shell deformation as a process of shell stiffness and bone strength. A cohort of 32 shells with two different wall thicknesses (3 and 4 mm) and 10(More)
The introduction of coatings on joint replacements was intended to reduce wear volumes, prevent corrosion and reduce metal ion release. However, retrieval analysis to confirm their in vivo performance has been limited. The aim of study was to examine the coating on a retrieved hip prosthesis to determine whether substantial damage or wear had occurred. A(More)