Learn More
The breakage of a stainless steel tracheostomy tube resulted in the death of a 19-year old man. Technical investigations identified corrosion as the mechanism of degradation of the braced joint between the plate and the external tube. Examination of a stock of 112 sterling silver tubes showed corrosive attacks on four tubes. Recommendations are presented(More)