Two hundred and sixteen Lightspeed instruments were evaluated microscopically for the presence of corrosion, surface debris, and alloy defects. The instruments were assessed morphometrically for consistency of physical design and dimensions by measuring and analyzing eight parameters of the instrument pilot tips, heads, and shafts. Results from visual(More)
Revolutions to separation and maximum torque at failure of 216 Lightspeed instruments were determined in an instron using a clockwise rotation. After instruments failed, the distance the instrument separated from the tip was measured. Comparison of the results with existing ANSI/ADA specification no. 28 showed that the Lightspeed far exceeded the values of(More)