The scatter to primary ratio (SPR) was measured on a scanning multislit full-field digital mammography system for different thickness of breast equivalent material and different tube voltages. Scatter within the detector was measured separately and was found to be the major source of scatter in the assembly. Measured total SPRs below 6% are reported for(More)
An evaluation of the dependence of detective quantum efficiency (DQE) on the incident energy spectrum has been made for mammography. The DQE dependence on the energy spectrum has been evaluated for energy-integrating detectors, photon-counting detectors, and detectors that measure the energy of each photon. To isolate the effect of the x-ray energy spectrum(More)
The physical performance of a scanning multislit full field digital mammography system was determined using basic image quality parameters. The system employs a direct detection detector comprised of linear silicon strip sensors in an edge-on geometry connected to photon counting electronics. The pixel size is 50 microm and the field of view 24 x 26 cm2.(More)
Silicon saw-tooth refractive lenses have been in successful use for vertical focusing and collimation of high-energy X-rays (50-100 keV) at the 1-ID undulator beamline of the Advanced Photon Source. In addition to presenting an effectively parabolic thickness profile, as required for aberration-free refractive optics, these devices allow high transmission(More)
A Fresnel-like X-ray lens can be constructed by a triangular array of identical prisms whose base corresponds to the 2pi-shift length. Each column of prisms is progressively shifted from the optical axis by an arbitrary fraction of the prism height. Similarly to the multi-prism lens, quasi-parabolic profiles are formed by a superposition of straight-line(More)
PURPOSE Spectral imaging is a method in medical x-ray imaging to extract information about the object constituents by the material-specific energy dependence of x-ray attenuation. The authors have investigated a photon-counting spectral imaging system with two energy bins for contrast-enhanced mammography. System optimization and the potential benefit(More)
In x-ray imaging, contrast information content varies with photon energy. It is, therefore, possible to improve image quality by weighting photons according to energy. We have implemented and evaluated so-called energy weighting on a commercially available spectral photon-counting mammography system. The technique was evaluated using computer simulations,(More)
We present an experimental and theoretical evaluation of an x-ray energy filter based on the chromatic properties of a prism-array lens (PAL). It is intended for small-scale applications such as medical imaging. The PAL approximates a Fresnel lens and allows for high efficiency compared to filters based on ordinary refractive lenses, however at the cost of(More)
Antiscatter grids are used in digital mammography to reduce the scattered radiation from the breast and improve image contrast. They are, however, imperfect and lead to partial absorption of primary radiation, as well as failing to absorb all scattered radiation. Nevertheless, the general consensus has been that antiscatter grids improve image quality for(More)