Tauhe authors would like to report a mechanical failure of a Thompson's prosthesis, 28 years post-implantation. A detailed examination of the specimen revealed no defects in the prosthesis and a dominating 'brittle component' fracture of the stem. In this context the detailed fractographic study by scanning electron microscopy (SEM) revealed no detrimental(More)